# COD calculator (Open reflux) in mg/L | NovaLIMS

> Free COD calculator by the open reflux method (IS 3025 (Part 58)). Enter your readings to get chemical oxygen demand in mg/L, with the formula, a worked example and typo checks.

Source: https://novalims.in/tools/calculators/cod

## How COD is calculated

Free COD calculator (open reflux method), following IS 3025 (Part 58). Enter 4 readings to get chemical oxygen demand in mg/L, with the formula filled in step by step.

1.  ### Chemical oxygen demand · mg/L
    
    `COD = (A − B) × N × 8 × 1000 / V`
    

## What you need

| Symbol | Reading | Unit | Notes |
| --- | --- | --- | --- |
| V | Volume of sample taken | ml | usually 1–50 ml |
| N | Normality of FAS | N | usually 0.01–0.5 N |
| A | FAS used for blank | ml | usually 0–50 ml (a 50 ml burette) |
| B | FAS used for sample | ml | usually 0–50 ml (a 50 ml burette) |

### Fixed values in the formula

| Symbol | Constant | Value |
| --- | --- | --- |
| 8 | Equivalent weight of oxygen | 8 |

## Worked example

Readings:

-   Volume of sample taken: **V = 20 ml**
-   Normality of FAS: **N = 0.25 N**
-   FAS used for blank: **A = 10.4 ml**
-   FAS used for sample: **B = 6.2 ml**

1.  ### Chemical oxygen demand
    
    `COD = (10.4 − 6.2) × 0.25 × 8 × 1000 / 20 = 420.0 mg/L`
    

Result: Chemical oxygen demand **420.0 mg/L**.

## Checks and units

-   **Typical values:** Domestic sewage: about 250–800 mg/L
-   The blank titre (A) should be larger than the sample titre (B). Check you haven't swapped them.
-   **Other units:** ppm (1 mg/L ≈ 1 ppm in dilute water).

## Questions

### How do you calculate COD by the open reflux method?

Chemical oxygen demand = (A − B) × N × 8 × 1000 / V. Here V is the volume of sample taken (ml), N is the normality of FAS (N), A is the FAS used for blank (ml), B is the FAS used for sample (ml). The result is in mg/L.

### What unit is COD reported in?

Chemical oxygen demand is reported in mg/L, to 1 decimal place. The calculator can also show it in ppm (1 mg/L ≈ 1 ppm in dilute water).

### Which standard does this COD calculation follow?

It follows the calculation in IS 3025 (Part 58). Check it against the current edition of the standard and your lab’s SOP before reporting.

### Can you show a worked example?

With volume of sample taken = 20 ml, normality of FAS = 0.25 N, FAS used for blank = 10.4 ml, FAS used for sample = 6.2 ml, the result is chemical oxygen demand = 420.0 mg/L.

### What is a typical COD value?

Domestic sewage: about 250–800 mg/L

### What mistakes should I look out for?

The blank titre (A) should be larger than the sample titre (B). Check you haven't swapped them.

### Is this COD calculator free?

Yes. It’s free with no sign-up. The calculation runs in your browser and nothing you type is sent to NovaLIMS.

## Related calculators

-   [Dissolved oxygen – Winkler (iodometric): Oxygen demand](https://novalims.in/tools/calculators/dissolved-oxygen)

[All water & waste water calculators →](https://novalims.in/tools/calculators/water)

Last reviewed 10 November 2026. Version 1. Follows IS 3025 (Part 58). [How we check our formulas](https://novalims.in/tools/calculators/methodology)

Check results against your own method and SOP before reporting. Found an error? [Tell us](mailto:hello@novalims.in?subject=Calculator%20check) and we’ll check it.