# Lead calculator (AAS / ICP on the filter digest) in µg/m³ | NovaLIMS

> Free lead calculator by the AAS / ICP on the filter digest method (IS 5182 (Part 22)). Enter your readings to get lead in PM10 in µg/m³, with the formula, a worked example and typo checks.

Source: https://novalims.in/tools/calculators/lead-ambient

## How lead is calculated

Free lead calculator (AAS / ICP on the filter digest method), following IS 5182 (Part 22). Enter 7 readings to get lead in PM10 in µg/m³, with the formula filled in step by step.

1.  ### Volume of air sampled · m³
    
    `V = Q × t`
    
2.  ### Lead in PM10 · µg/m³
    
    `C = (Mₛ − M_b) × Vₛ × Fₐ / (V × Fₜ)`
    

## What you need

| Symbol | Reading | Unit | Notes |
| --- | --- | --- | --- |
| Q | Sampling rate | m³/min | usually 0.5–2 m³/min |
| t | Sampling time | min | usually 60–1440 min |
| Mₛ | Concentration in the sample digest | µg/ml | usually 0–10 µg/ml |
| M_b | Concentration in the blank digest | µg/ml | usually 0–1 µg/ml |
| Vₛ | Total volume of extract | ml | usually 5–100 ml |
| Fₐ | Total exposed area of the filter | cm² | usually 10–600 cm² |
| Fₜ | Area of filter taken for digestion | cm² | usually 1–600 cm² |

## Worked example

Readings:

-   Sampling rate: **Q = 1.1 m³/min**
-   Sampling time: **t = 1440 min**
-   Concentration in the sample digest: **Mₛ = 0.042 µg/ml**
-   Concentration in the blank digest: **M\_b = 0.004 µg/ml**
-   Total volume of extract: **Vₛ = 25 ml**
-   Total exposed area of the filter: **Fₐ = 406 cm²**
-   Area of filter taken for digestion: **Fₜ = 50.8 cm²**

1.  ### Volume of air sampled
    
    `V = 1.1 × 1440 = 1,584.0 m³`
    
2.  ### Lead in PM10
    
    `C = (0.042 − 0.004) × 25 × 406 / (1,584.0 × 50.8) = 0.005 µg/m³`
    

Result: Lead in PM10 **0.005 µg/m³**.

## Checks and units

-   The sample digest reads lower than the blank.

## Questions

### How do you calculate lead by the AAS / ICP on the filter digest method?

Work through 2 steps in order. Volume of air sampled: V = Q × t. Lead in PM10: C = (Mₛ − M\_b) × Vₛ × Fₐ / (V × Fₜ). Here Q is the sampling rate (m³/min), t is the sampling time (min), Mₛ is the concentration in the sample digest (µg/ml), M\_b is the concentration in the blank digest (µg/ml), Vₛ is the total volume of extract (ml), Fₐ is the total exposed area of the filter (cm²), Fₜ is the area of filter taken for digestion (cm²). The result is in µg/m³.

### What unit is lead reported in?

Lead in PM10 is reported in µg/m³, to 3 decimal places.

### Which standard does this lead calculation follow?

It follows the calculation in IS 5182 (Part 22). Check it against the current edition of the standard and your lab’s SOP before reporting.

### Can you show a worked example?

With sampling rate = 1.1 m³/min, sampling time = 1440 min, concentration in the sample digest = 0.042 µg/ml, concentration in the blank digest = 0.004 µg/ml, total volume of extract = 25 ml, total exposed area of the filter = 406 cm², area of filter taken for digestion = 50.8 cm², the result is lead in PM10 = 0.005 µg/m³.

### What mistakes should I look out for?

The sample digest reads lower than the blank.

### Is this lead calculator free?

Yes. It’s free with no sign-up. The calculation runs in your browser and nothing you type is sent to NovaLIMS.

## Related calculators

-   [Arsenic and nickel (AAS / ICP on the filter digest): Metals & organics](https://novalims.in/tools/calculators/arsenic-nickel-ambient)
-   [Benzo(a)pyrene (Solvent extraction and GC): Metals & organics](https://novalims.in/tools/calculators/bap-ambient)

[All ambient air calculators →](https://novalims.in/tools/calculators/ambient-air)

Last reviewed 10 November 2026. Version 1. Follows IS 5182 (Part 22). [How we check our formulas](https://novalims.in/tools/calculators/methodology)

Check results against your own method and SOP before reporting. Found an error? [Tell us](mailto:hello@novalims.in?subject=Calculator%20check) and we’ll check it.

## Run every calculation inside your LIMS

NovaLIMS puts these calculators right beside each sample’s parameters, for every analyst in the lab.

[Book a Demo](https://novalims.in/contact) [Start Free Trial](https://novalims.in/contact?intent=trial)