Skip to content
690 nmPO₄³⁻

Phosphate

Stannous chlorideResult in mg/LIS 3025 (Part 31)

Turns blue · read at 690 nm

Formula
PO₄³⁻ =(A−B)×G.F×D

Your readings

5 of 5 filled

Example values shown.

ml
Abs
Abs
mg/L per Abs
Phosphate as PO₄³⁻Calculated live as you type
0.800 mg/L
Worked out with the formula above.

Calculations you finish this week are counted here, on this device only.

How phosphate is calculated

Free phosphate calculator (stannous chloride method), following IS 3025 (Part 31). Enter 5 readings to get phosphate as PO₄³⁻ in mg/L, with the formula filled in step by step.

  1. Phosphate as PO₄³⁻ · mg/L

    PO₄³⁻ = (A − B) × G.F × D

What you need

SymbolReadingUnitNotes
VVolume of sample takenmlusually 1–1000 ml
BAbsorbance of blankAbsusually 0–0.2 Abs
AAbsorbance of sampleAbsusually 0–2 Abs (the usual 0–2 absorbance range)
G.FGraph factormg/L per Absusually 0.01–500 mg/L per Abs
DDilution factor—usually 1–1000

Worked example

Readings:

  • Volume of sample taken: V = 50 ml
  • Absorbance of blank: B = 0.006 Abs
  • Absorbance of sample: A = 0.264 Abs
  • Graph factor: G.F = 3.1 mg/L per Abs
  • Dilution factor: D = 1
  1. Phosphate as PO₄³⁻

    PO₄³⁻ = (0.264 − 0.006) × 3.1 × 1 = 0.800 mg/L

Result: Phosphate as PO₄³⁻ 0.800 mg/L.

Checks and units

  • The sample reading (A) is lower than the blank (B), so the result would be negative. Check both.
  • Other units: ppm (1 mg/L ≈ 1 ppm in dilute water).

Questions

How do you calculate phosphate by the stannous chloride method?

Phosphate as PO₄³⁻ = (A − B) × G.F × D. Here V is the volume of sample taken (ml), B is the absorbance of blank (Abs), A is the absorbance of sample (Abs), G.F is the graph factor (mg/L per Abs), D is the dilution factor. The result is in mg/L.

What unit is phosphate reported in?

Phosphate as PO₄³⁻ is reported in mg/L, to 3 decimal places. The calculator can also show it in ppm (1 mg/L ≈ 1 ppm in dilute water).

Which standard does this phosphate calculation follow?

It follows the calculation in IS 3025 (Part 31). Check it against the current edition of the standard and your lab’s SOP before reporting.

Can you show a worked example?

With volume of sample taken = 50 ml, absorbance of blank = 0.006 Abs, absorbance of sample = 0.264 Abs, graph factor = 3.1 mg/L per Abs, dilution factor = 1, the result is phosphate as PO₄³⁻ = 0.800 mg/L.

What mistakes should I look out for?

The sample reading (A) is lower than the blank (B), so the result would be negative. Check both.

Is this phosphate calculator free?

Yes. It’s free with no sign-up. The calculation runs in your browser and nothing you type is sent to NovaLIMS.

Related calculators

All water & waste water calculators →

Check results against your own method and SOP before reporting. Found an error? Tell us and we’ll check it.

Run every calculation inside your LIMS

NovaLIMS puts these calculators right beside each sample’s parameters, for every analyst in the lab.