Skip to content
540 nmNO₂

Nitrogen dioxide

Modified Jacobs and Hochheiser (by shift)Result in µg/m³IS 5182 (Part 6)

Turns pink · read at 540 nm

Formulas, in order
Volume of air sampled · L · each shift
Vair =Q×t
NO₂ for the shift · µg/m³ · each shift
C =(Aₛ−A_b)×G.F×1000×DF×Vₛ(Vair×V×0.82)
Average nitrogen dioxide · µg/m³
NO₂ =avg(C)

Your readings

4 of 4 filled

Example values shown.

µg per Abs
ml
Sampling efficiencyConstant · fixed by the method0.82🔒
ShifttSampling timeminQFlow rateL/minVₛFinal volume of absorbing solutionmlAₛAbsorbance of sampleAbsA_bAbsorbance of blankAbsVairVolume of air sampledLCNO₂ for the shiftµg/m³
Shift 124017.6
Shift 224020
Shift 324015.9
Shift 424014
Shift 524012.9
Shift 624014.8
Average nitrogen dioxideCalculated live as you type
15.9 µg/m³µg/m³
Within the usual range. 24-hour standard (NAAQS): 80 µg/m³.

Calculations you finish this week are counted here, on this device only.

How nitrogen dioxide is calculated

Free nitrogen dioxide calculator (modified Jacobs and Hochheiser (by shift) method), following IS 5182 (Part 6). Enter 8 readings to get average nitrogen dioxide in µg/m³, with the formula filled in step by step.

  1. Volume of air sampled · L · for each shift

    Vair = Q × t

  2. NO₂ for the shift · µg/m³ · for each shift

    C = (Aₛ − A_b) × G.F × 1000 × DF × Vₛ / (Vair × V × 0.82)

  3. Average nitrogen dioxide · µg/m³

    NO₂ = avg(C)

What you need

SymbolReadingUnitNotes
G.FGraph factorµg per Absusually 0.5–100 µg per Abs
DFDilution factor—usually 1–100
VAliquot taken for analysismlusually 1–50 ml
tSampling timeminone per shift; usually 10–600 min
QFlow rateL/minone per shift; usually 0.1–5 L/min
VₛFinal volume of absorbing solutionmlone per shift; usually 5–100 ml
AₛAbsorbance of sampleAbsone per shift; usually 0–2 Abs
A_bAbsorbance of blankAbsone per shift; usually 0–0.2 Abs

Fixed values in the formula

SymbolConstantValue
0.82Sampling efficiency0.82

Worked example

Readings:

  • Graph factor: G.F = 18 µg per Abs
  • Dilution factor: DF = 1
  • Aliquot taken for analysis: V = 10 ml
  • Sampling time: t = 240, 240, 240, 240, 240, 240 min
  • Flow rate: Q = 1, 1, 1, 1, 1, 1 L/min
  • Final volume of absorbing solution: Vₛ = 30, 30, 30, 30, 30, 30 ml
  • Absorbance of sample: Aₛ = 0.072, 0.081, 0.066, 0.059, 0.055, 0.062 Abs
  • Absorbance of blank: A_b = 0.008, 0.008, 0.008, 0.008, 0.008, 0.008 Abs
  1. Volume of air sampled

    Vair = 240.0, 240.0, 240.0, 240.0, 240.0, 240.0 L

  2. NO₂ for the shift

    C = 17.6, 20.0, 15.9, 14.0, 12.9, 14.8 µg/m³

  3. Average nitrogen dioxide

    NO₂ = avg(17.6, 20.0, 15.9, 14.0, 12.9, 14.8) = 15.9 µg/m³

Result: Average nitrogen dioxide 15.9 µg/m³.

Checks and units

  • Typical values: 24-hour standard (NAAQS): 80 µg/m³

Questions

How do you calculate nitrogen dioxide by the modified Jacobs and Hochheiser (by shift) method?

Work through 3 steps in order. Volume of air sampled: Vair = Q × t. NO₂ for the shift: C = (Aₛ − A_b) × G.F × 1000 × DF × Vₛ / (Vair × V × 0.82). Average nitrogen dioxide: NO₂ = avg(C). Here G.F is the graph factor (µg per Abs), DF is the dilution factor, V is the aliquot taken for analysis (ml). The result is in µg/m³.

What unit is nitrogen dioxide reported in?

Average nitrogen dioxide is reported in µg/m³, to 1 decimal place.

Which standard does this nitrogen dioxide calculation follow?

It follows the calculation in IS 5182 (Part 6). Check it against the current edition of the standard and your lab’s SOP before reporting.

Can you show a worked example?

With graph factor = 18 µg per Abs, dilution factor = 1, aliquot taken for analysis = 10 ml, sampling time = 240, 240, 240, 240, 240, 240 min, flow rate = 1, 1, 1, 1, 1, 1 L/min, final volume of absorbing solution = 30, 30, 30, 30, 30, 30 ml, absorbance of sample = 0.072, 0.081, 0.066, 0.059, 0.055, 0.062 Abs, absorbance of blank = 0.008, 0.008, 0.008, 0.008, 0.008, 0.008 Abs, the result is average nitrogen dioxide = 15.9 µg/m³.

What is a typical nitrogen dioxide value?

24-hour standard (NAAQS): 80 µg/m³

Is this nitrogen dioxide calculator free?

Yes. It’s free with no sign-up. The calculation runs in your browser and nothing you type is sent to NovaLIMS.

Related calculators

All ambient air calculators →

Check results against your own method and SOP before reporting. Found an error? Tell us and we’ll check it.

Run every calculation inside your LIMS

NovaLIMS puts these calculators right beside each sample’s parameters, for every analyst in the lab.